FLUKE 17B Digital MultiMeter
Accuracy is specified for 1 year after calibration, at operating temperatures of 18 �C to 28 �C, relative humidity at 0 % to 75 %. Accuracy specifications take the form of: �([% of Reading] + [Number of Least Significant Digits])
Function Range Resolution Accuracy Model 15B Model 17B 3.0 % + 3 1.0 % + 3 1.0 % + 10 0.5 % + 3 1% typical 5 Function Range Resolution Accuracy Model 15B Model 17B 1.5 % + 3 1.5 % + 3 1.5 % + 3 1.5 % + 3 1.5 % + 3 1.5 % + 3 1.5 % + 3 1.5 % + 3 1.5 % + 3 1.5 % + 3 1.5 % + 3 1.5 % + 3 Function Input Impedance (Nominal) Normal Mode Rejection >10 M� <100 pF >10 M� <100 pF Thw whole package includes: 1) New FLUKE 17B Multimeter, 2) Thermocouple; 3) A pair of test leads; 4) English manual;
To take a look at the detail English Manual:http://assets.fluke.com.cn/evtmanuals/15b17b__cmeng0000.pdf Customer Support Phone Number: (408) 329-7985
AC Volts (40 to 500 Hz) v
400.0 mV 1
0.1 mV
3.0 % + 3
4.000 V 40.00 V 400.0 V 1000 V
0.001 V 0.01 V 0.1 V 1 V
1.0 % + 3
DC Millivolts M
400.0 mV
0.1 mV
1.0 % + 10
DC Volts V
4.000 V 40.00 V 400.0 V 1000 V
0.001 V 0.01 V 0.1 V 1 V
0.5 % + 3
Diode Test
1.000 V
0.001 V
10 %
Temperature 3 �C (K-type thermocouple)
50 �C to 400 �C 0 �C to 50 �C -55 �C to 0 �C
0.1 �C
NA
2 % + 1 �C � 2 �C 9 % + 2 �C
Resistance (Ohms)
400.0 �4.000 k�40.00 k�400.0 k�4.000 M�40.00 M�
0.1 �0.001 k�0.01 k�0.1 k�0.001 M�0.01 M�
0.5 % + 3 0.5 % + 2 0.5 % + 2 0.5 % + 2 0.5 % + 2 1.5 % + 3
0.5 % + 3 0.5 % + 2 0.5 % + 2 0.5 % + 2 0.5 % + 2 1.5 % + 3
Capacitance 4 c
50.00 nF 500.0 nF 5.000 �F 50.00 �F 100.0 �F
0.01 nF 0.1 nF 0.001 �F 0.01 �F 0.1 �F
2 % + 5 2 % + 5 5 % + 5 5 % + 5 5 % + 5
2 % + 5 2 % + 5 5 % + 5 5 % + 5 5 % + 5
Frequency Hz (10 Hz � 100 kHz)
50.00 Hz 500.0 Hz 5.000 kHz 50.00 kHz 100.0 kHz
0.01 Hz 0.1 Hz 0.001 kHz 0.01 kHz 0.1 kHz
NA
0.1 % + 3
Duty Cycle
0.1 % to 99.9 %
0.1%
NA
1. Manual Range only 2. Diode test open circuit test voltage is 1.1 V to 1.6 V and short circuit current is < 0.6 mA (typical). 3. Temperature specifications do not include thermocouple errors. After inserting the thermocouple plug into the meter, allow several minutes for thermal stabilization. 4. Specifications do not include errors due to test lead capacitance and capacitance floor (may be up to 1.5 nF in the 50 nF range). For the 17B, errors can be reduced by using the relative feature. 5. For values between 10% and 90% duty cycle at 50 Hz.
AC Current (40 to 200 Hz)
400.0 �A
0.1 �A
4000 �A
1 �A
AC Current (40 to 200 Hz)
40.00 mA
0.01 mA
400.0 mA
0.1 mA
AC Current (40 to 200 Hz)
4.000 A6 10.00 A
0.01 A 0.01 A
DC Current
400.0 �A
0.1 �A
4000 �A
1 �A
DC Current
40.00 mA
0.01 mA
400.0 mA
0.1 mA
DC Current
4.000 A6
0.01 A
10.00 A
0.01 A
6. When in the 4A range, display will show 4000 counts, please ignore the last digit.
Overload Protection
Common Mode Rejection Ratio
AC Volts v
1000 V
>60 dB at dc, 50 or 60 Hz
_
DC Volts V
1000 V
>100 dB at dc, 50 or 60 Hz
>45 dB at 50 or 60 Hz
106 V Hz Max
Customer Support Skype: Aidetech
| Maximum Voltage between any Terminal and Earth Ground: | 1000 V |
| W Fuse Protection for mA �A inputs: | 500 mA, 1000 V FAST, Min. Interrupt Rating 20,000 A |
| W Fuse Protection for A input: | 10 A, 1000 V FAST, Min. Interrupt Rating 20,000 A |
| Display: | Digital: 4000 count updates 3/sec |
| Temperature: | Operating: 0 �C to 40 �C, Storage: -30 �C to 60 �C indefinitely (to -40 �C for 100 hours) |
| Operating Altitude: | 0 to 2000 meters |
| Temperature Coefficient: | 0.1 X (specified accuracy)/ �C (<18 �C or >28 �C) |
| Electromagnetic Compatibility: | Complies with FCC Part 15, Class B, IEC 61326, 3 V/m, performance criterion B |
| Rated Transient Overvoltage: | 4 kV (1.2 x 50 �s) Peak for measurement Categories I, II, and III. |
| Relative Humidity: | Noncondensing < 10 �C 90% from 10 �C to 30 �C; 75% from 30 �C to 40 �C |
| Relative Humidity, 40 M� Range: | 80% from 10 �C to 30 �C; 70% from 30 �C to 40 �C |
| Battery Type: | 2 X AA, NEDA 15A, IEC LR6 |
| Battery Life: | Alkaline: 500 Hours |
| Size (H x W x L): | 180 mm x 89 mm x 51.5 mm (with holster) |
| Weight: | 425 grams |
| Certifications: | CMC, P |
| Safety Compliance: | IEC 61010-1, 2000 CAT I 1000 V, CAT II 600 V, and CAT III 300 V overvoltage standards |
| Overvoltage installation categories per IEC 61010-1, 2000: The Meter is designed to protect against transients in these catagories: | |
| CAT I: From high-voltage low-energy sources, e.g., electronic circuits or a copy machine. CAT II: From equipment supplied from the fixed installation, e.g., Televisions, PCs, portable tools and household appliances. CAT III: From equipment in fixed equipment installations, e.g., installation panels, feeders and short branch circuits, and lighting systems in large buildings. | |
Tell us what you think about this item, share your opinion with other people. Please make sure that your review focus on this item. All the reviews are moderated and will be reviewed within two business days. Inappropriate reviews will not be posted.